Description
Endo Activator Tips are used with all Sonic Endo Activator or Dentsply Endoactivator. The Tips have an easy snap on/snap off design. They are strong and flexible made up with medical grade polymer. They are very easy to
differentiate from other tips as the Large tips are Blue 35/04 (Large), Medium tips are red 25/04 (Medium), Small tips are yellow 15/02 (Small)
Directions For Use
API Sonic Endo activator Tips Can be very easily placed on the API sonic Activator as it has a snap
On / Snap-off design.
- First, prepare the canal to produce a fully tapered shape
- Fill The Pulp Chamber with EDTA, NaOCl or any other final Rinsing solution
- Selection of the irrigation tip that manually fits loosely within 2 mm of working length is must
- There are Three Types of irrigation tip available – Small, Medium and large
- Place the barrier sleeve over the complete length of the handpiece
- Attach the Irrigator tip over the barrier-protected handpiece
- The Irrigator tip should snap
on firmly, promoting a secure connection with the handpiece
- The Irrigator tip is then passively inserted in the canal within 2mm of the WL
- The irrigator tips are moved in short 2-3 mm vertical strokes
- Hydrodynamically agitate the intracanal solution for 30-30 seconds Irrigate
- Then use intracanal suction to eliminate loose debris
- Repeat the above steps for each intracanal irrigant used.
Irrigator Tip Removal:
- Firmly Support the head of the handpiece with your thumb while grasping the white circular
the flange portion of the irrigator tip with opposite thumb and forefinger
- Use your thumb to maintain the support of the handpiece and pull the irrigator tip STRAIGHT off.
Keep the handpiece head and irrigator tip aligned During the removal process
- Keep the handpiece head and irrigator tip aligned During the removal process
- Use your thumb to maintain the support of the handpiece and pull the irrigator tip STRAIGHT
off
- The irrigator tip SNAP ON / OFF in exactly the same alignment.